Figure 6.49. Select Normalized Reading Screen.
7.
Touch <Accept> to return to the Manual Decay Test screen.
8.
Touch <Select Normalized Reading>. The Select Normalized Reading screen is
displayed.
Figure 6.50. The Manual Decay Test Screen with Test Ended message and data
displayed.
The last reading taken is highlighted as the normalized reading. To select a
different normalized reading, touch to highlight the reading and touch
<Accept>. The new reading is selected and will serve as the normalized
reading and the system returns to the manual decay test screen.
NOTE:
The user can select the normalized reading while the decay test is
underway. If a normalized reading has not been selected before selecting <End Test>,
the system automatically displays the Select Normalized Reading screen from
which the user can select a reading to use as normalized.
9.
Touch <Save>.
10.
If the appropriate staff member is not displayed, touch <Staff>. Touch any line on
the display to highlight the staff member. Touch <Accept> to return to the Save Results
screen.
10.
Touch <Accept> to select the staff member, save the data, and return to the Manual
Decay Test screen.
11.
Touch <New Test> to perform the test again.