Dose Calculator Status Displays
The following table contains a list of the Status Displays for Dose Calibrators:
Table B.2. Dose Calculator Status Display Table
Display
Description
Warming Up
The detector is in a power warm-up state.
Initializing
Count down:
•
Initial background reading
•
A calculation is in effect.
Background PASS
No Reported Problems – no errors in effect, ready for measuring activity
WARNING n-
<description> {1 error code}
WARNING n-n-n
Two or more error codes may be present.
FAIL: n-
<description>
FAIL: n-n-n
Two or more error codes may be present.
Warnings are errors flagged by the self-tests, but not considered permanent or serious enough
to cause a failure.
Failures (FAIL) are considered serious enough to not use the detector for any activity
measurements. The detector will automatically be taken offline by the system.