8.
Specifications
All-In-One Flat Panel Display:
10.8" Color Touch-Screen, Windows 10 Operating System and USB
connectivity.
Dimensions:
11" w x .375" depth x 7" (28 x 1 x 18 cm)
Line Voltage:
100 to 240 VAC, auto selectable by the power supply
Line Frequency:
50/60 Hz, detachable line cord, built-in EMI filter and transient suppression
Auxiliary Port:
Two USB ports, one RS-232 port
Isotope Selection Keys:
Twelve pre-programmed – Tc-99m, Tl-201, I-123, I-131, Cs-137, Co-
57, Xe-133, Ga-67, In-111, F-18, Y-90s, Ba-133 and a full alphabetical list of 97 isotopes.
Activity Range:
0.01 µCi to 100 Ci (.0004 MBq to 3700 GBq) of Tc-99m or 25 Ci of F-18
Energy Range:
25 keV to 3 MeV photons
Response Time:
One to two seconds for doses greater than 200 µCi; three seconds for doses
greater than 20 µCi; 50-100 seconds below 20 µCi of Tc-99m with default threshold; threshold
adjustable to reduce counting time
Electrometer Linearity:
± 1% or 0.2 µCi, whichever is greater, up to 40 curies of Tc-99m, ±
1.5% up to 100 curies of Tc-99m
Electrometer Accuracy:
± 1% or 0.2 µCi, whichever is greater
Overall Accuracy:
± 3% or 0.3 µCi, whichever is greater; overall accuracy is affected by such
factors as the accuracy of the specific source calibration, geometric variations due to sample
volume or configuration, detector linearity, electrometer accuracy and readout accuracy
Repeatability:
± 0.3% above 1 mCi short term (24 hr); 1% long term (one yr)
Detector:
Well-type pressurized ionization chamber, with Argon fill gas
Detector Shielding:
.25" (6.3
mm) lead on all sides except top well opening; supplementary shielding available
Detector Linearity:
± 1% or 0.2 µCi, whichever is greater
Detector Unit:
Dimensions:
6" dia x 15.5" h (15.24 x 39.37 cm)
Well I.D.:
2.75" dia x 10.5" h (7 x 26.7 cm)
Well I.D. with Liner:
2.5" dia x 10.25" h (6.35 x 26 cm)
Lead Shielding:
.25" lead (6.3 mm)
Weight:
35 lb (16 kg)
Power Requirements:
100 to 240 VAC, 0.38 – 0.15 amps, auto switching
Certification:
ETL
listed to UL 60601-1 and to CAN/CSA C22.2 No. 601-1-M90 IEC 60601-1, IEC 60601-1-4 and
IEC 60601-1-2 and CE marked
Warranty:
Two years parts and labor.