NOTE:
Make sure the correct tube combination has been selected, as highlighted on the
display.
7.
Touch <Save>. If the appropriate staff member is not displayed, touch <Staff>.
Touch any line on the display to highlight the staff member. Touch <Accept> to
return to the Save Results screen.
8.
<New Test> to perform the test again.
9.
Touch <Accept> to select the staff member, save the data, and return to the Calicheck
Test screen.
10.
Touch <Save Data> to save the data.
11.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
12.
Touch <Back> to return to the Lineator Testing screen.
Linearity Test History
Since there are several acceptable methods for performing linearity testing, the Linearity
History allows the user to select the category of test for which the history is to be
displayed. The following sub-screen is displayed when Linearity Test History is selected:
Figure 6.79. Linearity History Screen.
Select an option by touching the appropriate button. Each test category is described in detail
below.
Decay Test History
The View Decay Test data screen displays the decay tests performed on the dose calibrator.
The display lists individual decay tests with the most recent test as the default. Use the <
▲
>
and <
▼
> to scroll to view the test results desired. Any test or reading outside the
acceptable variance is shown in red. The user can view the decay plot for the test being
shown on the display.