Figure 7.3.
The Moly Assay Zero Background reading.
To Perform A Moly Assay:
1.
At the Home screen, touch <Nuclear Pharmacy>. The Nuclear Pharmacy screen is displayed
(see Figure 7.1).
2.
Touch <Moly Assay>. The Moly Assay Zero Background Reading screen is displayed (see
Figure 7.3).
3.
The Moly Assay Background screen allows the user to change the background counting
time. The screen displays the current background time. To change the background
counting time, select <Change Background Time>. A message will be displayed stating,
“Changing Background, Time will be Retained and have a Global Effect Throughout the
System.”
Figure 7.4.
Change Background Time Message.
Select <Accept>.
4.
The Set Count Background Time screen is displayed.