Select Current Isotope
Figure 4.20. The
Select Current Isotope Screen.0
To Select Any Isotope From The Full Isotope Lists:
1.
From the <Home> screen, select <Utilities>. The Utilities screen is displayed.
2.
Select <Dose Calibrator Configuration>. The Dose Calibrator Configuration screen is
displayed.
3.
Select <Isotope List>.
4.
The Isotope List screen is displayed.
5.
Touch <Select Current Isotope>. The Select Current Isotope screen is displayed.
6.
Use the <
▲
> or <
▼
> keys to scroll through the list until the desired isotope is displayed.
7.
Touch the desired isotope to highlight the selection.
Figure 4.21. Isotope Selection Example
8.
Touch <Accept> to select the desired isotope and return to the Isotope List screen with the
new isotope selected, or touch <Cancel> to return to the Isotope List screen without
making a new selection.
The selected isotope is displayed on the Measure screen list as illustrated below.