NOTE:
Touch <Pause> at any time during the automated portion to pause the test
and allow use of the dose calibrator for counting and other functions, such as
choosing <Measure> to return to standard DOSE CALIBRATOR OPERATION.
NOTE:
When in the manual portion of the test, the user can press <Measure> to
display the Measure screen and use the system as a Dose Calibrator. Touch
<Return to Decay Test> on the Measure screen to return to the automated decay
test. Touch <Resume Test> to resume the Automated Decay Test.
Figure 6.56. Dose Calibrator Measure Screen.
15.
Touch <End Test> to end the test.
Figure 6.57. The Semi-Automated Decay Test Ended Screen.
16.
Touch <Save> to save the data. The Save Results screen is displayed.
17.
If the appropriate staff member is not displayed, touch <Staff>. Touch any line on the
display to highlight the staff member. Touch <Accept> to return to the Save Results
screen.
18.
Touch <Accept> to select the staff member, save the data, and return to the Semi-
Automated Decay Test screen.
19.
Touch <New Test> to perform the test again.
20.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
21.
Touch <Back> to return to the Semi-Automated Decay Test screen.
22.
Touch <Back> a second time to return to the Linearity Testing screen.