desired regional format, and the UTC time zone format.
To Set Date/Time:
1.
Touch <Regional Date/Time Settings> on the System Configuration screen.
Figure 4.13. System Configuration Screen
2.
The Regional Set Date/Time Settings screen is displayed (see Figure 4.12).
3.
Touch the Region Format field to highlight it and use the scroll bar at the right to
navigate through the list until the desired region format is located. Select the region
format by touching the <Accept> key.
Figure 4.14. Region Format Selection
4.
Touch the Time Zone field to highlight it and use the scroll bar at the right to navigate
through the list until the desired time zone is located. Touch the time zone to
highlight and select it.