HELP
Same as the ? command. It displays all available commands with a brief description.
Available commands are:
Command
Description
BAUD #
Set the BAUD Rate, where:
•
1 = 19K
•
2 = 38K
•
3 = 56K
•
4 = 115K.
BKGND
Zero Background.
DCNT
Displays the number of connected detectors.
DS {1/0}
Data Streaming, where:
•
1 = ON
•
0 = OFF.
DSTAT {#}
Status for all detectors or the one specified.
DT
Display the Atomlab Date/Time.
ECHO {#}
Command Line Echo, where:
•
1 = ON
•
0 = OFF.
HELP
Displays the Help screen.
ISO {#}
Select Isotope index.
MODEL
Displays the Atomlab Model #.
PWD {1/0}
Sets the Password Override, where:
•
1 = ON
•
0 = OFF.
QADATA
Displays a summary of QA data usage.
REQDATA
Requests stored QA data matching the given criteria.
REQRECCNT
Requests a count of QA records matching the given criteria.
RESET
Resets/Restarts the Atomlab program.
STAT
Displays the full detector status.
VER
Displays the firmware versions.
ISO {#}{#}
Displays or sets the isotope assigned to a detector. For a single detector system, the second
parameter is not required. The first parameter is an isotope index number; 1 to 98 for the
default isotopes, 99 to 118 for custom isotopes. Refer to the separate isotope index document
at the end of this appendix for details. The second parameter is the detector number. If no
parameter is given, the current assigned isotope for each detector is displayed. Use this