Touch <Delete Isotope>.
•
The isotope is deleted from the list.
11.
When information has been entered into all of the fields, touch <Save>.
12.
The following message is displayed:
Figure 4.55. Saved Message
13.
Touch <Accept> to save the isotope(s) entered in the Expanded Constancy List.
14.
Touch <Back> to return to the Sealed Source Screen and verify that the new source is listed.
Note:
When an isotope(s) has been entered in the Expanded Constancy List, the system
marks the number entered in the Source ID field with an asterisk as illustrated below:
Figure 4.56. Sealed Source List Screen Displaying Isotope Entered in Sealed
Expanded Constancy List
To Edit A Sealed Source:
1.
At the Sealed Sources List screen, touch anywhere on the line to highlight the isotope to
edit.
2.
Touch <Edit>. The Sealed Sources Setup screen is displayed.