Figure 4.38. The Change Dial Values Screen
7.
Use the <
▲
> or <
▼
> keys to scroll through the list until the desired isotope is displayed.
8.
Touch the desired isotope to highlight the selection.
9.
Touch the <Change Dial Value> icon at the bottom of the screen to display the dial setting
entry screen.
10.
A keypad is displayed on which the user is to enter the new value.
Figure 4.39. Dial Setting Keypad
Use the <
▲
> or <
▼
> keys to change the value, or touch <Clear> to empty the field and
enter a new value using the keypad.
11.
Touch <Accept> to record the new dial value and return to the Change Dial Values screen
with the new dial value displayed in the New Dial Value field.
12.
Touch <Cancel> to exit and return to the Change Dial Values screen without saving any
changes.
13.
If desired, touch the following icons:
•
<Restore Selected Factory Dial Values> to remove the change(s) made.
•
<Restore Factory Dial Values> and <Accept> to revert to the factory defaults or touch
<No> to cancel restoring the factory defaults.
•
<Print> to print out the updated Isotope Library.