Calibration Value Display:
Up and down arrow control with dial values from 0.0 to 999.9.
Detector Type:
Detector:
The detector is a well-type pressurized ionization detector, with argon-filled gas. A
replaceable plastic liner is included.
Dimensions:
Well Opening:
2.75 in. (7 cm).
Well Depth:
10.5 in. (26.7 cm).
Well Opening with Liner:
2.5 in. (6.35 cm).
Well Depth with Liner:
10.25 in. (26 cm).
Detector Shield:
.25 in. (6 mm) lead surrounding ionization detector, with well opening.
Detector Bias:
355 +/- 5 volts.
Environmental Operating Conditions:
Temperature: 10 - 30°C.
Humidity:
0 to 90% rh, non-condensing.
For optimum performance, the Atomlab 500 should be operated in a normal laboratory
environment where the temperature and humidity are maintained for normal human comfort
and the ambient radiation level is low and constant.
Physical Data:
Table C.5. Atomlab 500 Size and Weight.
Display Unit:
Detector Unit:
Size:
11.81" x 8.07" x 0.39"
(300 x 205 x 9.9 mm)
6" dia x 15.5" h
(15.24 x 39.37 cm)
Weight:
1.94 lbs (0.87 kg)
38 lb (17.3 kg)
Power:
The system uses an XP power supply for medical use, model # pdm60us15 or the alternate ICC
Nexergy (Pei Genesis) model MWA065015A-11A (modified).
Line Voltage:
100 to 240 VAC, auto selectable by the power supply.
1.50 – 0.75 AMPS.
Line Frequency:
50/60 Hz.
Detachable line cord. Built-in EMI filter and transient suppression.
WARNING:
Turn off all system power before inserting or removing connectors.
AVERTISSEMENT:
Éteignez toute la puissance du système avant d'insérer ou de
retirer des connecteurs.