6.7.8 Spare parts
WARNING
Risk of injury if incorrect spare parts are used!
ü
Using incorrect or faulty spare parts poses a danger to
personnel and patients and can cause damage, mal-
functions, or total failure.
1. Use only original spare parts from the manufacturer or
spare parts approved by the manufacturer.
2. Contact the manufacturer with any doubts or concerns.
NOTICE
1. Failure to comply voids warranty if damage occurs
2. .Use of non-approved spare parts voids warranty.
Obtain spare parts through authorized dealers or directly from the manu-
facturer.
6.7.9 Securing device during maintenance
WARNING
Risk of death through unauthorized or accidental power-up!
ü
Unauthorized or accidental power-up of the device can
cause serious injury and even death during mainte-
nance.
1. Before turning the device back on, make sure that all
covers are installed and functional and that there are
no hazards present.
2. Always follow the procedure described below to secure
the device against accidental switch-on.
1. Disconnect the power cord and turn the device off at the main
switch.
2. Notify responsible parties about work on the device
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Safety
88