Materials:
¡
Use 5% aluminum acetate solution/white wine vinegar
1. Apply the 5% aluminum acetate solution/white wine vinegar to a lint-
free cloth.
2. Remove the spots with the cloth.
3. Rinse the surface with clean water.
4. Dry with a soft, lint-free cloth.
NOTICE
Visible impurities, such as blood and gross soiling/bioburden,
must always be thoroughly cleaned off prior to disinfection.
Materials:
¡
Quaternary ammonium compounds
NOTICE
Material damage from disinfectants such as lodophor!
Using disinfectants such as lodophor causes spotting on the OR
table.
1. Use only quaternary ammonium compounds for disinfec-
tion.
NOTICE
Material damage from sodium hydroxide and sodium hypochlor‐
ite.
The use of sodium hydroxide (lye) and sodium hypochlorite
(bleach) on coated aluminum parts can lead to surface changes
and affect the service life of the product.
1. Do not use sodium hydroxide or sodium hypochlorite sol-
utions.
Removing spots (e.g.,
blood/plaster)
Disinfection
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760