3.3 Attaching the pads
CAUTION
Risk of injury if incorrect pads are used!
ü
Using incorrect pads can compromise their antistatic
properties during radiation. This can cause an electro-
static shock that can seriously injure the patient and per-
sonnel and severely damage critical components.
1. Use only antistatic, conductive, and radiolucent pads.
2. The pads should be initially installed only by the manu-
facturer's technicians.
1. To attach the pads, align them flush with the edge of the table.
2. Press the pads down.
¡
The pads are held in place with Velcro strips.
NOTICE! Sample design. Pad shapes and variants can differ depending
on the scope of delivery.
NOTICE
Removing pads!
To remove pads, slide your hand between the pad and base sec‐
tion and unhook the Velcro fastener.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Installation and initial start-up