NOTICE
Grounding pads, straps and other components can be cleaned
immediately after soiling by washing them with a mild soap solu‐
tion. Remove stubborn spots with a mild, soapy solution and a
soft brush..
1. Use only quaternary ammonium compounds or phenol-containing
disinfectants for cleaning according the manufacturer's instructions.
4.17.1 Charging the batteries
The batteries need charging when the BATT indicator light (1) is continu-
ously lit red.
Personnel:
¡
Medical specialists
1. Connect the OR table to the power supply (“Connecting to the pow-
er supply” section)
4.18 Hand pendant care
This section provides instructions for care, cleaning, and disinfection of
BERCHTOLD OPERON® D-Series Surgical Table Hand Pendant.
1. If dropped, inspect immediately for damage, otherwise remove from
service until a thorough inspection for cracks and dam-aged seals
can be performed
2. When the hand control is not actively being used ensure it is hung
from the medical rail on the head section.
3. Prevent hand control from being exposed to gross soiling and bio-
burden.
4. If hand control is exposed to gross soiling/bio-burden during a proce-
dure disconnect hand control from the table immediately.
5. Remove gross soiling/bio-burden with a cloth dampened with a mild
detergent then dry.
6. After cleaning, ensure no fluids are allowed to pool or stand on the
hand control.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation
58