Function
Description
Lateral tilt (TILT)
The TILT function stops if any table
section nears the floor, base, or col-
umn, or if internal components are
at risk of colliding.
HEIGHT
The HEIGHT function stops if any ta-
ble section nears the floor, base, or
column.
Kidney elevator (optional), (KIDNEY) The KIDNEY function stops if the
back section is greater than or
equal to 20°.
Back plate (BACK)
The BACK function stops at an up-
ward angle of 20° if the kidney ele-
vator is in use (regardless of its
height).
LEG
The LEG function stops if the leg
section nears the floor, base, or col-
umn.
Longitudinal slide (D820, D850 only)
(SLIDE)
The SLIDE function stops if any table
section nears the floor, base, or col-
umn.
CHAIR, FLEX, and LEVEL preset posi-
tions
The automatic functions stop any
function that could come into con-
tact with the floor, base, or column.
The LEVEL function activates other
functions to prevent collisions and
finish leveling the tabletop.
NOTICE
Material damage due to incorrect movement using SPLIT LEG
function!
1. Split leg sections are not taken into account by the soft-
ware-controlled limits during movement. Movement
does not stop and can result in a collision and material
damage to the OR table.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation
28