4.2 Patient positioning
WARNING
Risk of injury due to incorrect patient positioning!
. The OR table can over balance if the patient is positioned incor‐
rectly, causing serious injury
1. Do not lean on the different sections or use them for sup-
port.
2. Comply with specifications provided in the technical da-
ta on patient weight, size and positioning.
3. At a Trendelenburg angle greater than 15 °, the maxi-
mum patient weight is 617 lb. (280 kg), and 500 lb. (227
kg) reverse orientation. This ensures that the longitudinal
slide will work correctly.
1. Position the patient on the OR table according to the technical data
(“Patient positioning” section).
4.3 Software-controlled limits of the OR table
CAUTION
Risk of pinching or crushing when operating the leg section
(LEG)!
1. There is a risk of pinching or crushing when the LEG func-
tion is activated.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation
26