Ipump Pain Management System Operator’s Manual
5-26
CD07-19-E4-766
Programming the Bolus Dose
The bolus dose is either programmed and delivered automatically at the beginning of therapy, or initiated by
the clinician during the course of therapy as a supplemental (clinician) bolus [see “Programming a
Supplemental (Clinician) Bolus,” 5-28]. If a programmed bolus dose has not been started, the pump will
display the programmed dosage instead of zeroes.
Action
Message Displayed
1.
Use the
⇐
and
⇒
keys and
↑
to program the bolus dose.
NOTE:
If no bolus is desired, enter zero.
0000.0 mg SET
↑
BOLUS
2.
If necessary, press the
Clear/Silence
key to reset the displayed value to zero.
3.
Press
Enter
when the desired value is displayed.
0100.0 mg SET
↑
BOLUS
4.
Connect the pump tubing set to the patient’s access device.
START BEGINS Rx
ENTER REVIEWS Rx
Oct02_0719E4766.book Page 26 Friday, August 3, 2007 1:46 PM