Ipump Pain Management System Operator’s Manual
5-29
CD07-19-E4-766
3.
At the
SELECT ACTION
prompt, use the
⇑
key to display the
SET BOLUS
prompt, then press
Enter
.
SELECT ACTION
SET BOLUS
4.
At the
SET BOLUS
prompt, use the
⇐
and
⇒
keys and
⇑
key to set the desired
bolus dose.
Press
Clear/Silence
to make the displayed value zero.
Press
Enter
when the desired value is displayed.
00.0 mL SET
↑
BOLUS
5.
Press
Enter
to review or change the prescription. See “Review the
Prescription,” 5-37 or “Changing the Prescription During Infusion,” 5-35.
OR
Press
START
to begin infusing the supplemental (clinician) bolus.
START BEGINS Rx
ENTER REVIEWS Rx
6.
If you pressed
START
, the pump resumes the infusion and displays
BOLUS INFUSING
. The bolus delivery will continue until the bolus dose is
delivered or
STOP
is pressed twice in 1 second.
Example:
BOLUS INFUSING
0001.3 mg
Action
Message Displayed