Ipump Pain Management System Operator’s Manual
10-4
CD07-19-E4-766
Table 10-2 Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the pump should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV contact
Floors should be wood, concrete, or ceramic tile. If
the floors are covered with synthetic material, the
relative humidity should be at least 30%.
±
8 kV air
±
8 kV air (1)
Functions normally except at PRINTER/COMM
connector pins.
±
15 kV air (2)
Pump may stop infusing and alarm at levels above
±
8kV air.
±
4 kV air (1)(2)
Avoid touching the PRINTER/COMM connector
pins directly.
Oct02_0719E4766.book Page 4 Friday, August 3, 2007 1:46 PM