Ipump Pain Management System Operator’s Manual
9-3
CD07-19-E4-766
Concentration Ranges
When units of mg are selected, the below values are limited to the same volumes as for
prescriptions entered in mL.
0.1 mg/mL to 99.9 mg/mL in 0.1 mg/mL increments
1µ g/mL to 999 µg/mL in 1 µg/mL increments
Security Features
Locking cover
3-digit programmable security code
Latched tubing door
Indicators
Alphanumeric description via LCD display
Red Alert light
Green Infusing light
Audible tones: Minimum audible alarm sound level: 50 dBa
Battery
9-volt alkaline
Drive Mechanism
DC Motor, microprocessor-controlled, precision linear peristaltic pumping mechanism
Printer Port
1200 Baud, 8 data bits, no parity and 1 stop bit
Housing
Shock- and vibration-resistant Acrylonitrile Butadiene Styrene (ABS)
Size
12.4 cm x 8.6 cm x 4.6 cm (4.9" x 3.4" x 1.8") without cover
Component
Description
Oct02_0719E4766.book Page 3 Friday, August 3, 2007 1:46 PM