Ipump Pain Management System Operator’s Manual
7-5
CD07-19-E4-766
Transporting and Storing the Pump
When unpacked, store the pump without the battery in a clean and dry environment to safeguard against
prolonged exposure to dust and moisture. This storage area should meet the following environmental
guidelines:
Q
Temperature range: -20ºC to 60ºC (-4ºF to 140ºF)
Q
Relative humidity: 10% to 95% (non-condensing)
If conditions fall outside these limits, Baxter recommends that the device be repackaged in the original
shipping materials.
Note:
When storing the pump, remove the 9-volt battery from the pump. Always store the pump with the tubing door closed.
Functional Testing
Perform as required, but recommended every 6 months.
Check
Action
Entire device
Schedule functional test by qualified biomedical personnel or authorized service personnel as specified in the
Ipump Pain Management System Service Manual.