Ipump Pain Management System Operator’s Manual
10-7
CD07-19-E4-766
Notes for Table 10-2
Note 1
! WARNING !
The pins of the PRINTER/COMM connector should not be touched. The PRINTER/COMM
connector should not have objects inserted into it other than the mating connector. ESD
precautionary procedures should be used when inserting the mating connector into the
connector.
It is recommended that all staff involved in the use or servicing of this device receive training in the
ESD precautionary procedures. The ESD precautionary procedures should include:
• an introduction to the physics of electrostatic charge;
• the voltage levels that can occur in normal practice;
• the damage that can be done to electronic components by electrostatic discharge;
• an explanation of methods to prevent build up of electrostatic charge;
• an explanation on why to discharge one's body to earth or a large metal object;
• an explanation on bonding oneself by means of a wrist strap to earth prior to servicing.
Note 2
The pump was tested to the EMC requirements of IEC 60601-2-24: 1998. IEC 60601-2-24, the
particular standard for infusion pumps, requires higher test levels.
Note 3
Input/output lines are exempt because they are less than 3.0 meters long.
Note 4
The pump automatically transfers to battery operation if there is a loss of mains power.
Table 10-2 Guidance and Manufacturer's Declaration – Electromagnetic Immunity — continued