Ipump Pain Management System Operator’s Manual
6-7
CD07-19-E4-766
PCA BUTTON
NOT CONNECTED
Situation: A PCA button is required to continue the pump’s operation.
Action: Connect the PCA button to the pump.
NOTE:
If the pump is configured to use the
START
key to administer a PCA dose, this
message will not occur.
PREVENTIVE
MAINTENANCE DUE
Situation: The configured preventive maintenance period has elapsed.
Action: Perform the preventive maintenance procedures as described in “Preventive
Maintenance,” 7-1.
NOTE:
After its initial occurrence, the
PREVENTIVE MAINTENANCE DUE
message will
appear each time the pump is turned on
⎯
until the preventive maintenance is reset.
PUMP LEFT IN
CONFIG MODE
Situation: The configuration mode time period has elapsed.
Action: Press
Enter
to cancel the alert and restart the configuration mode time period.
PUMP LEFT IN
PROGRAMMING MODE
Situation: The programming mode time period has elapsed.
Action: Press
Enter
to cancel the alert and restart the programming mode time period.
Note:
The pump retains all prescription data entered prior to the timeout.
Table 6-1
Alert Messages and Responses — continued
Alert Message
Situation/Action