Ipump Pain Management System Operator’s Manual
6-14
CD07-19-E4-766
EMPTY
Situation: The fluid bag is empty.
Action:
1.
Replace the fluid bag with a filled fluid bag.
2.
Reprogram the fluid volume and prime as described in “Set the Fluid Volume in
the Reservoir,” 5-15 and “Prime the Tubing Set,” 5-16.
RELEASE PCA
BUTTON
Situation: The PCA button has been pressed continuously for 6 minutes.
Action:
1.
Release the PCA button.
2.
Advise the patient to press the PCA button briefly when requesting a PCA dose.
NOTE:
If this alarm occurs and the PCA button is not being pressed intentionally, the PCA
button may have a mechanical or electronic fault. Do not use the pump. Take the pump
out of service and contact your nearest authorized service center.
Table 6-2
Alarm Messages and Responses — continued
Alarm Message
Situation/Action
Oct02_0719E4766.book Page 14 Friday, August 3, 2007 1:46 PM