Ipump Pain Management System Operator’s Manual
3-5
CD07-19-E4-766
Q
If the pump is battery-operated and the infusion has been started, stop the pump by pressing
STOP
twice
in one second and replace the battery (the pump turns off during the battery change). Then, turn the
pump on. The pump will retain the prescription and therapy history.
To resume the infusion, press
START
at the
PUMP READY PRESS START
or
PUMP READY START
OR CLEAR
screen after the power-on self test runs. The infusion will resume after the memory test
and upstream occlusion test (if the pump is configured for upstream occlusion detection) are
complete.
If the pump is configured as CODE ONLY, the
SELECT ACTION
options may be accessed prior to
resuming the infusion; press
Clear/Silence
at the
PUMP READY START OR CLEAR
screen after the
power-on self test runs. See “Restarting the Infusion,” 5-33, for more information.
Note:
Always follow manufacturer’s recommendations and applicable local regulations when using or disposing of batteries.
Connecting the AC Adapter
The AC adapter must be used to plug the pump into an AC electrical power source. The 9-volt alkaline
battery must be inserted in the battery compartment as a backup power source in the event of AC power
interruption and to allow patient ambulation.