Ipump Pain Management System Operator’s Manual
10-6
CD07-19-E4-766
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(>95% dip
in U
T
) for 5 sec.
<5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(>95% dip
in U
T
) for 5 sec.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the pump requires continued operation during
power mains interruptions, it is recommended that
the pump be powered from an uninterruptible
power supply or a battery.
User should always have battery installed per
operator’s manual. (4)
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic characteristic of a
typical location in a typical commercial or hospital
environment.
400 A/m (2)
Pump may stop infusing and alarm at levels above
3A/m.
Table 10-2 Guidance and Manufacturer's Declaration – Electromagnetic Immunity — continued