Ipump Pain Management System Operator’s Manual
10-5
CD07-19-E4-766
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrical fast
transient burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for input/
output lines
±
2 kV for power
supply lines
Not Applicable (3)
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common
mode
±
1 kV differential
mode
±
2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Table 10-2 Guidance and Manufacturer's Declaration – Electromagnetic Immunity — continued