Ipump Pain Management System Operator’s Manual
10-9
CD07-19-E4-766
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
3 V/m (3)
10V/m(3)
d=4
√
P 80 MHz to 800 MHz
d=7.7
√
P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
b
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
c
, should be less than the
compliance level in each frequency range.
d
Interference may occur in the vicinity of equipment marked
with the following symbol:
Pump may stop infusing and alarm at levels above 3V/m.
Table 10-3 Guidance and Manufacturer's Declaration –
Electromagnetic Immunity- for Life-Supporting Equipment and Systems — continued