Ipump Pain Management System Operator’s Manual
5-14
CD07-19-E4-766
Set the Concentration
The units displayed on the pump’s LCD are determined by the selected programming unit and mode. If the
current (initial) prescription settings are being used, that concentration setting is displayed instead of zeroes.
If mL units are selected, you do not have to enter a concentration.
Action
Message Displayed
1.
Use the
⇐
or
⇒
keys and
⇑
key to set the correct concentration.
00.0 mg/mL SET
↑
CONC.
2.
If you make a mistake, press the:
⇑
key until the correct value appears.
OR
⇐
or
⇒
key to reposition the
↑
.
OR
Clear/Silence
key to reset the previous value to 0.
000 μg/mL SET
↑
CONC.
3.
Press
Enter
when the desired value is displayed.
Example:
10.0 mg/mL SET
↑
CONC.
100 μg/mL SET
↑
CONC.