Ipump Pain Management System Operator’s Manual
3-4
CD07-19-E4-766
Change the Battery
The 9-volt alkaline battery should be changed regularly. If battery voltage drops below the required level:
Q
A
LOW BATTERY
alert will appear on the pump’s screen,
Q
The red
Alert
and green
Infusing
LEDs will flash, and
Q
An audible alert will sound.
You must do one of the following before changing the battery as described in “Install the 9-volt Alkaline
Battery,” 3-2:
Q
If the pump is powered by the AC adapter, you can replace the battery at any time without interrupting
operation. During the battery change, the pump will display the
PCA
message and the
BATTERY MISSING
alert, but will not interrupt service.
Q
If the pump is battery-operated and the infusion has not been started, press
Enter
to acknowledge the
alert and replace the battery (the pump turns off during the battery change). Then, turn the pump on and
re-enter the prescription after the power-on self test as described in “Using the Pump,” 5-1.
Oct02_0719E4766.book Page 4 Friday, August 3, 2007 1:46 PM