Ipump Pain Management System Operator’s Manual
5-44
CD07-19-E4-766
7.
As you scroll, the following messages are displayed:
7.1
The date and time of the start of the shift. (A new shift is started
whenever the operator clears the shift totals.)
Example:
SHIFT STARTED
03/17/99 12:00AM
7.2
For PCA doses only, the total injections (
TOTAL INJ
) administered
and total attempts (
TOTAL ATT
) recorded per shift.
Example:
0004 TOTAL INJ
0004 TOTAL ATT
7.3
The Shift Total for any PCA, Basal+PCA, Continuous, or bolus
infusions.
Example:
0031.5 mL
SHIFT TOTAL
8.
After you view the SHIFT TOTAL screen, you may (as an option) choose to:
8.1
Press
Clear/Silence
to clear the totals and begin a new shift.
8.2
Wait for the
INITIAL SETTINGS
of the prescription to appear on the
screen.
SHIFT TOTALS
CLEARED
NOTE:
Pressing
Clear/Silence
will reset the total infused and the “injections versus attempts” to zero and set the time and
date of the new shift to the time/date that the pump was cleared.
Action
Message Displayed