Special conditions for safe use
• Consult the manufacturer if dimensional information on
the flameproof joints is necessary.
• Follow the manufacturer’s instructions to reduce the
potential of an electrostatic charging hazard.
• Consult the manufacturer for genuine replacement flange
fasteners. M10x35 hexagon socket cap screws of
• ISO 12.9 DIN912 grade steel (zinc-plated) or better, with
a minimum yield strength of 135,000 psi are acceptable
alternatives.
• The sensor body temperature code rating is dependent
on the process temperature (see the table below). It is
assumed that the external surface of the sensor body will
be, at worst case, equal to the process temperature. In all
cases, the electronics assembly will be marked T6, as it will
be locally mounted for process temperatures up to 85°C
and will be remotely mounted for process temperatures
exceeding 85°C.
NOTE:
The remote mount electronics option is not yet
available for the
Sentinel LCT8
model.
Mounting
Process
Temperature (TP)
T-code
Local
TP ≤ 85°C
T6
Remote
TP ≤ 85°C
T6
Remote
85°C < TP ≤ 100°C
T5
Remote
100°C < TP ≤ 120°C
T4A
(For USA and Canada only)
Remote
100°C < TP ≤ 135°C
T4
Remote
135°C < TP ≤ 140°C
140°C
124
Summary of Contents for Panametrics Sentinel LCT8
Page 1: ...Sentinel LCT8 User s manual...
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 30: ...20...
Page 98: ...88...
Page 122: ...no content intended for this page 112...
Page 132: ...122...