3
Safe handling and preparation
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Read, follow, and keep the instructions for use.
►
Use the product only in accordance with its intended use, see Intended
use.
►
Remove the transport packaging and clean the new product, either
manually or mechanically, prior to its initial sterilization.
►
Store any new or unused products in a dry, clean, and safe place.
►
Prior to each use, inspect the product for loose, bent, broken, cracked,
worn, or fractured components.
►
Do not use the product if it is damaged or defective. Set aside the prod-
uct if it is damaged.
►
Replace any damaged components immediately with original spare
parts.
►
To avoid damage to the working end: Carefully insert the product
through the working channel (e.g. trocar).
Single-use products
The product is gamma-sterilized and supplied in sterile packaging.
The product must not be reused.
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Read, follow, and keep the instructions for use.
►
Use the product only in accordance with its intended use, see Intended
use.
►
Do not use products from open or damaged sterile packaging.
►
Prior to each use, inspect the product for loose, bent, broken, cracked
or fractured components.
►
Do not use the product if it is damaged or defective. Set aside the prod-
uct if it is damaged.
►
Do not use the product after its use-by date.
Safe operation
Art. no.
Designation
FW086SU
Drill
∅
2.9 mm for smooth-shaft screws
FW088SU
Favored Angle screw drill
∅
2.9 mm for
∅
4 mm
screws
FW662SU
S
4
C1/C2 drill
∅
2.4 mm
FW666SU
Drill
∅
2.4 mm
FW667SU
Drill
∅
2.9 mm
DANGER
Infection hazard for patients and/or users and
impairment of product functionality due to reuse.
Risk of injury, illness or death due to contamination
and/or impaired functionality of the product!
►
Do not reprocess the product.
WARNING
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
the product.
WARNING
Risk of injury to patients due to damaged instru-
ments!
S
4
C instruments are extremely precise and very
sensitive.
►
Handle S
4
C instruments with the greatest of
care.
►
Check that dropped or damaged S
4
C instruments
are accurately calibrated, or send them to the
Aesculap Technical Service.
WARNING
Risk of injury to patients due to interruption of the
navigation!
►
Prior to the operation, plan the configuration of
the operating room, the assembly of the instru-
ments and the alignment of the reference star.
►
Ensure that the navigation camera has an unre-
stricted view of the reflective marker spheres of
the instruments.
WARNING
Risk of injury to patients due to imprecise instru-
ments!
►
Ensure that the instruments used are not bent
or damaged.
►
Before use, check the precision of the instru-
ments, particularly that of fine instruments. For
this, hold the instrument tip in the pivot point
of the Brainlab instrument calibration matrix
version 4.