An electrical problem has arisen with solenoid valve EV4, probably a short circuit.
Check if CN9 connector is properly inserted in the motherboard and if there are no uncovered wires causing
a short circuit;
Check if the solenoid valve coil is sound and is not "burnt". Replace the solenoid valve if this is the case
(refer to the instructions in 6.2.2).
E60: Mos Sol.Valve 5
An electrical problem has arisen with solenoid valve EV5, probably a short circuit.
Check if CN9 connector is properly inserted in the motherboard and if there are no uncovered wires causing
a short circuit;
Check if the solenoid valve coil is sound and is not "burnt". Replace the solenoid valve if this is the case
(refer to the instructions in 6.2.2).
E61: Mos Sol.Valve 6
An electrical problem has arisen with solenoid valve EV6, probably a short circuit.
Check if CN9 connector is properly inserted in the motherboard and if there are no uncovered wires causing
a short circuit;
Check if the solenoid valve coil is sound and is not "burnt". Replace the solenoid valve if this is the case
(refer to the instructions in 6.2.2).
E62: Mos Sol.Valve 7
An electrical problem has arisen with solenoid valve EV7, probably a short circuit.
Check if CN11 connector is properly inserted in the motherboard and if there are no uncovered wires causing
a short circuit;
Check if the solenoid valve coil is sound and is not "burnt". Replace the solenoid valve if this is the case
(refer to the instructions in 6.2.2).
E63: Mos Sol.Valve 8
An electrical problem has arisen with solenoid valve EV8, probably a short circuit.
Check if CN11 connector is properly inserted in the motherboard and if there are no uncovered wires causing
a short circuit;
Check if the solenoid valve coil is sound and is not "burnt". Replace the solenoid valve if this is the case
(refer to the instructions in 6.2.2).
E64: Mos ACS Fan
An electrical problem has arisen with fan 1, probably a short circuit.
Check if the fan supply connector is properly inserted in the motherboard and if there are no uncovered wires
causing a short circuit;
Check if the fan motor is sound and is not "burnt". Replace the fan if this is the case.
E65: Mos ACS Pump
An electrical problem has arisen with fan 2, probably a short circuit.
Check if the fan supply connector is properly inserted in the motherboard and if there are no uncovered wires
causing a short circuit;
Check if the fan motor is sound and is not "burnt". Replace the fan if this is the case.
E66: Mos Door
An electrical problem has arisen with the actuator that drives the automatic door (for models with automatic
doors only), probably a short circuit.
Check if the actuator supply connector C12 is properly inserted in the motherboard and if there are no
uncovered wires causing a short circuit;
Check if the actuator is sound and is not "burnt". Replace the actuator if this is the case.