1-8
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
A
3. The obtained specific gravity value of the sample is corrected using correction coefficients
a
and
b
.
SGx = a × SGx´ + b
SGx:
Sample’s specific gravity after calibration correction
SGx´:
Sample’s specific gravity before calibration correction
SG
H
:
Specific gravity reference of standard solution (high)
SG
L
:
Specific gravity reference of standard solution (low)
K
H
:
Specific gravity of standard solution (high)
K
L
:
Specific gravity of standard solution (low)
Furthermore, high concentrations of glucose (GLU) and protein (PRO) will affect specific gravity.
Therefore, specific gravity is corrected using the concentrations of glucose and protein obtained
from test strip measurement.
SG = SGx – C
GLU
– C
PRO
SG:
Specific gravity after GLU and PRO correction
SGx:
Specific gravity obtained from formula 3.
C
GLU
:
Corrected GLU value
C
PRO
:
Corrected PRO value
Correction values (Unit of concentration: mg/dL)
GLU concentration
0
10
30
50
70
100
Correction value
0.0000
0.0000
0.0001
0.0002
0.0003
0.0004
150
200
300
500
1000
OVER
0.0006
0.0008
0.0012
0.0020
0.0040
0.0040
PRO concentration
0
5
10
20
30
50
70
Correction value
0.0000
0.0000
0.0000
0.0001
0.0001
0.0002
0.0002
100
150
200
250
300
600
OVER
0.0003
0.0005
0.0006
0.0008
0.0009
0.0018
0.0018
Color-Tone Measurement
An internal turbidity/color measurement unit measures sample color using transmitted light.
470 nm/525 nm/635 nm
multiple wavelength LED
Cylindrical cell (i.d. 1.8 mm)
Photodiode (transmitted light)
Photodiode (Scattered light)