AX-4030
5-3
AUTION MAX AX-4030 OPERATING MANUAL
T221
A/D overflow
Cause
. The transport tray is not installed correctly.
2. The waste box is overflowing with used test strips that are blocking the measurement light
and the optical unit movement.
3. The test strip was not placed correctly in the optical unit.
4. The white plate is dirty.
Remedy . Press
OK
to clear the trouble.
↓
Install the transport tray correctly (see page 4-9).
2. Press
OK
to clear the trouble.
↓
Pull out the waste box and discard the used test strips (see page 4-3).
↓
Remove the test strips or other obstacles that are blocking the measurement light and the
optical unit movement.
3. Press
OK
to clear the trouble.
↓
Install the transport tray correctly.
↓
Remove the obstacles that are blocking the measurement light and the optical unit
movement.
↓
Remove the test strips scattered on the analysis section.
4. Press
OK
to clear the trouble.
↓
Replace the white plate with a new one (see page 4-32), and perform a check measurement
(see page 2-43).
T222
Optical unit trouble
Cause
. The white plate is dirty.
2. The test strip was not transported correctly.
Remedy . Press
OK
to clear the trouble.
↓
Replace the white plate with a new one (see page 4-32), and perform a check measurement
(see page 2-43).
2. Press
OK
to clear the trouble.
↓
Remove the test strips scattered on the analysis section.