5-6
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
T270
Measurement lever drive trouble
Cause
. A sample rack was forced into the sampler.
2. The measurement- or return-side lever of the sampler did not work correctly.
Remedy . Press
OK
to clear the trouble.
↓
Take the sample racks out of the sampler.
2. Press
OK
to clear the trouble.
↓
Check the sampler. If the sampler is damaged, turn off the power and contact your distributor.
T271
Return lever drive trouble
Cause
. A sample rack was forced into the sampler.
2. The measurement- or return-side lever of the sampler did not work correctly.
Remedy . Press
OK
to clear the trouble.
↓
Take the sample racks out of the sampler.
2. Press
OK
to clear the trouble.
↓
Check the sampler. If the sampler is damaged, turn off the power and contact your distributor.
T272
Loading side full
Cause
Another rack was added to the loading side when the sampler was fully loaded with 5 racks.
Remedy Press
OK
to clear the trouble.
↓
Take the sample racks out of the loading side of the sampler.
T280
Overflow
Cause
An overflow has occurred, or the tubing is detached.
Remedy Press
OK
to clear the trouble.
↓
Check the drain pinch valves. Replace the damaged tubes with new ones (see page 4-30).
↓
Check to see if tubing is disconnected.
T290
Cover open
Cause
The front cover, maintenance cover, side cover, or feeder protective cover is open.
Remedy Press
OK
to clear the trouble.
↓
Ensure all the covers are closed.