1-38
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
2
Disconnect the liquid sensor cord and tube of the washing solution bottle.
1
Disconnect the liquid sensor cord from the washing solution sensor terminal “S” on the rear
panel.
2
Disconnect the tube of the washing bottle from the washing solution joint “W” on the rear panel.
3
Disconnect cables to the optional part and external devices.
1
If an optional hand-held barcode reader or external device is connected to the instrument,
disconnect their cables from the instrument.
1.6.6
Detaching the Sampler
Prepare: Phillips screwdriver
1
Detach the sampler.
1
Carefully lift the sampler straight upward with both
hands and pull to the front.
•
This unhooks the sampler in 2 locations.
Tube
Liquid sensor cord
❷
❶
To washing solution bottle
W
D
H
S
Hook