2-46
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
•
When check measurement is completed, the
obtained result will be reported on the screen.
5
Press
to print the result.
6
Press
Go back
to return to the
[Maintenance]
screen.
6
Measure the second test strip (gray strip).
1
Press
3
to go to the
[Check measurement]
screen.
2
Select the same feeder as the one for the first strip.
•
For detailed instructions, see step
3
.
3
Take a gray test strip out of the container.
4
Load the gray test strip into the feeder selected in
step
2
.
•
For detailed instructions, see step
5
.
Check measurement
(5300) 2006-01-09 14:11
Measurement completed.
***** 2006-01-09 14:10 ****
430[nm] 89.3 %
500[nm] 93.1 %
565[nm] 88.5 %
635[nm] 9.7 %
760[nm] 100.0 %
Print Go back
Reflectivity
Wavelengths
Maintenance
(5000) 2006-01-09 14:15
1
S.G. cell washout
2
Clean washing bath and tray
3
Check measurement
4
Maintenance information
Check measurement
(5300) 2006-01-09 14:15
Feeder No.
[
1
]
OK Go back
Gray check strip
Loading slot
Black marker
Black marker facing down
Gray check strip