Press
Start
to start check measurement.
•
When check measurement is completed, the
obtained result will be reported on the screen.
6
Press
to print the result.
7
Evaluate the check measurement results.
1
Check the two printed result reports to see if the obtained reflectivity of each wavelength falls
within the range listed on the container label.
REFERENCE: For information on the result reports of check measurement, see
“2.8.6 Check
Measurement Result Report”
on page 2-56.
When the results are:
Within range
The instrument is working normally. Go to step
9
.
Out of range
Either the instrument or the test strips are abnormal. Perform a check measurement again. Press
Go back
to return to the
[Maintenance]
screen. Take the remaining two check strips out of the
container, and repeat from step
2
-
4
on page 2-44. Go to step
8
when completed.
NOTE:
If the printed results report includes an error message, clear the error as follows.
Reflection light intensity drift: Internal measurement operation is interfered by lights coming
from outside of the instrument. Remove the cause, and use the
same check strips to retry check measurement.
Test strip out-of-position:
The check strip is out of position. Use the same check strips to
retry check measurement.
Start Go back
Check measurement
(5300) 2006-01-09 14:16
Measurement completed.
***** 2006-01-09 14:15 ****
430[nm] 89.1 %
500[nm] 93.5 %
565[nm] 89.0 %
635[nm] 9.4 %
760[nm] 100.0 %
Print Go back
Reflectivity