AX-4030
1-39
AUTION MAX AX-4030 OPERATING MANUAL
2
Disconnect the sampler cables from the instrument.
1
Loosen the earth anchoring screw with a Phillips
screwdriver and disconnect the earth line from the
instrument.
2
Insert the tip of a Phillips screwdriver in the screw
holes on the splash guard to loosen the two screws
inside.
3
Slide the splash guard upward.
4
Temporarily tighten one of the screws using the
screwdriver to prevent the splash guard from
dropping.
5
Disconnect the two sampler cables from the
connectors on the lower part of the front panel.
•
The sampler is now detached completely from the
instrument.
1.6.7
Relocating the Instrument
1
Move the instrument to its new location.
IMPORTANT: For safety reasons, always move the instrument with the help of at least one
other person. Hold the bottom of the instrument with both hands when carrying it.
2
See
“1.4 Installation”
on page 1-18 to install the instrument.
Earth line
Earth
anchoring
screw
Screw hole
Splash guard
Cable
Sampler connectors