2-32
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
2
Set the measurement parameters.
If you use the test strips in the main feeder, skip this step and go to step
3
. The main feeder is
indicated by a light blue background on the standby screen.
1
To use the test strips in the non-main feeder, press
to go to the
[STAT reservation]
screen.
2
Below
[Sample feeding]
, choose
[Use rack]
and
press
OK
to go to the next screen.
3
Below
[Test strip]
, select the feeder that contains the
test strips to use.
Below
[Measurement No.]
, select the measurement
start number for STAT measurement.
•
For instructions, see step
4
in
“2.4.2 Port STAT
Measurement: Measuring a Higher Priority
Sample”
on page 2-27.
4
Press
OK
.
3
Load the STAT and control rack into the sampler.
1
Place the STAT and control rack on the loading side
of the sampler.
•
If there are some racks waiting for normal
measurement on the loading side, keep the rack at
the aspirating position (furthest from you) as it is
and slide the other racks toward you to make room
for the STAT and control rack.
•
If the loading side is full with racks, remove one of
them and insert the STAT and control rack.
During normal measurement, the aspirating nozzle descends into the ports
of the rack located furthest from you. For your safety, do not move the rack
in the process of aspiration.
NOTE:
If samples spill on the sampler, immediately wipe them off. Crystallized samples may
hinder smooth transportation of the sample racks, and consequently cause trouble.
STAT reservation
Sample feeding
[
Use rack
]
STAT reservation
Use rack
Test strip
[
1: 9EB
]
Measurement No.
<0001>
OK Cancel
Rack in aspirating
STAT and control rack