AX-4030
5-2
AUTION MAX AX-4030 OPERATING MANUAL
Measurement error
Description A measurement error occurred.
Cause
. Measurement was not made on this sample.
* This error occurs when the instrument is configured to skip measurement when a barcode
is misread.
2. Light intensity varied during measurement.
3. The reflectivity of a measurement items exceeded 20%.
4. The test strip is out of the correct position.
5. Sampling failed.
Remedy
. See
“Skipped”
on page 5-7.
2. See
“Reflection light intensity drift”
on page 5-7.
3. See
“Excess reflectivity”
on page 5-8.
4. See
“Test strip out-of-position”
on page 5-8.
5. See
“Abnormal sampling”
on page 5-8.
Abnormal data
Description Abnormal colors were developed on one or more reagent pads on the test strip.
Cause
Abnormal colored urine was measured, or the obtained result includes abnormal data.
Remedy
Check the sample.
Positive sample
Description The obtained result includes positive values for one or more measurement items.
Cause
The instrument measured a positive sample
Remedy
Check the sample.
S.G. measurement error
Description The instrument could not obtain a correct S.G. calibration result.
Cause
. No S.G. measurement result is present.
2. The S.G. measurement result exceeds the specified range.
3. The S.G. cell is dirty.
4. Abnormally colored urine was measured.
Remedy
and 2. Perform S.G. calibration (see page 2-34).
3. Wash the S.G. cell (see page 4-24).
4. Check the sample.
Turbidity measurement error
Description The instrument could not obtain a correct turbidity measurement result.
Cause
. No turbidity measurement result is present.
2. An abnormal turbidity measurement result was obtained.
Remedy
and 2. Turn off the power, and contact your distributor.