2.4.3
Rack STAT Measurement:
Measuring Higher Priority Samples
Rack STAT measurement allows you to have up to 7 samples measured prior to the samples
waiting for normal measurement in the sampler. This is accomplished by loading the higher priority
samples into the STAT and control rack and placing the rack between the sample racks currently
loaded in the sampler. Rack STAT measurement can also be carried out during standby.
Prepare: STAT and control rack, sample tubes, and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
1
Prepare samples for STAT measurement.
1
Pour samples for STAT measurement in sample
tubes.
•
See
“2.3.5 Preparing Samples”
on page 2-21.
2
Place the sample tubes straight into ports 1 to 7 with
the orange adapters of the STAT and control rack.
REFERENCE: Leave ports 8 to 0 with the dark
blue adapters empty as they are for control
measurement.
Orange adapters
Keep empty.
STAT and control rack
Black label