AX-4030
4-3
AUTION MAX AX-4030 OPERATING MANUAL
D
3
Replace the tubes.
1
Unhook the tubes.
2
Disconnect both ends of the tube from the joints.
•
One end should be removed from a T-joint.
3
Connect both ends of a new tube to the joints.
4
Feed the tube through the backward hook.
NOTE:
The drain pinch
valve has two hooks.
The tube must pass
through the backward
hook.
4
Close the covers.
1
Close the side cover (a), feeder protective cover (b),
and maintenance cover (c) in that order.
NOTE:
After replacing the tubes, check trouble
“T280” does not occur at the next startup. If it
occurs, check the bottom of the instrument and
drain tube connections for leaks. If there is a leak,
secure the connections.
Replaceable tubes
T-joint
Backward hook
Feeder
protective
cover
(a)
(b)
(c)
Side
cover
Maintenance cover