AX-4030
2-35
AUTION MAX AX-4030 OPERATING MANUAL
D
4
Load the S.G. solutions into the sample rack.
1
Transfer at least 2 mL of high solution into a sample
tube. Ensure the liquid level is at least 40 mm above
the bottom of the tube.
IMPORTANT: An insufficient amount of high
solution may cause incorrect calibration
results.
2
Place the low solution tube into port 1 of the sample
rack, and the high solution tube into port 2.
IMPORTANT: Load the two solutions into the
correct ports respectively. Loading them into
the wrong ports may cause an error.
5
Load the sample rack into the sampler.
1
Hold the sample rack with the marker facing you
and place it on the loading side of the sampler.
•
First, place the rack at the frontward of the loading
side, fit the recesses at the lower right of the rack
into the guides inside the sampler, and slide the rack
backward.
NOTE:
If solution spills on the sampler, immediately
wipe it off. Crystallized solutions may hinder
smooth transportation of the sample racks, and
consequently cause trouble.
Min. 2 mL water
High solution
Min. 40 mm
Port 1: low solution
Port 2: high solution
Marker