2-36
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
B
2.5.2
Starting S.G. Calibration
1
Access the S.G. calibration parameters.
1
On the standby screen, press
S.G. Cal
to go to the
[S.G. calibration]
screen.
2
Set the obtained standard S.G. values.
Prepare the notes you took in step
3
in
“2.5.1 Preparing S.G. Standard Solutions”
on page 2-34.
1
Below
[Standard solution (Low)]
, enter the S.G.
value of the low solution.
•
Use the numeric keys (0 to 9).
•
The S.G. value acceptable here is between 1.000 to
1.050.
REFERENCE: For example, to enter “.030”,
press
1
,
0
,
3
,
0
in that order. A decimal point is
automatically added. To correct the entry, press
BS
to delete the digits one by one, and enter
a new value.
2
Press
to proceed to the
[Standard solution
(High)]
entry field.
3
Below
[Standard solution (High)]
, enter the S.G.
value of the high solution.
•
Use the numeric keys (0 to 9).
•
The S.G. value acceptable here is between 1.000 to
1.050.
4
Press
OK
to save your entries.
•
The next
[S.G. Calibration]
screen will appear.
Ready for measurement
2006-01-09 10:30
Measurement start No.
0001
Test strips in feeder 1
1: 9EB
Test strips in feeder 2
2: 9EB
Menu Meas No. Strip S.G. Cal
S.G. Calibration
Standard solution (Low)
<1.00
0
>
Standard solution (High)
<1.040>
BS OK Cancel
S.G. Calibration
Standard solution (Low)
<1.002>
Standard solution (High)
<1.04
0
>
BS OK Cancel