User and Programming Manual
XDS Series
53
7.2 Routine
Calibration
Equipment Setup:
Connect the required test equipment to the power supply as shown in Figure 7-1. The DMM can be
connected to the common output bus bar at the rear of the unit.
Never connect the load wire to the SENSE terminal at the Rear Panel terminal strip. The load must
be connected to the POWER bus bars. The external sense input of the DC supply must be
connected at the same point as the DMM used to measure voltage. The DMM used to measure
current must be connected across the reference shunt. Make sure the LO side of both DMM's is
connected to the negative bus bar.
L
R
SHUNT
CURRENT
VOLTAGE
DC SUPPLY
(rear view)
POS.
NEG.
Measure output
voltage at sense
connection.
Figure 7-1: Test Equipment Hookup for Routine Output and Measurement Calibration